Main features
> Touch key board 6 keys.
> Direct readout on a graphic display.
> Data displayed:
Selected speed...................................................................................................r.p.m.
Selected spindle......................................................................................................SP
Viscosity reading...............................................................................cP (mPa·s) or cSt
Percentage of full scale............................................................................................%
Sample temperature....................................................................................... ℃ or ℉
Shear Rate (with coaxial spindles) ...................................................................SR (s-1)
Shear Stress (with coaxial spindles) .............................................................SS (N/m2 )
Density (introduced by the user) .......................................................................g/cm3
> Viscosity reading: dynamic viscosity
(cP or mPa·s) or kinematics viscosity (cSt).
> Unit converter SI to CGS.
> Program features:
Time to torque: target torque pre-setting device.
Time to stop: target time pre-setting device.
9 working memories.
> AUTO-TEST with sound and visual malfunction alarm.
> AUTO-RANGE function.
> Temperature reading by PT100 (optional).
> User-enabled viscosity and temperature (optional) calibration.
> 7 language options.
> USB-A (PT100, USB Stick, PC)
Technical specifications
> Precision: ± 1% of full scale
> Resolution:
With low viscosity adapter: 0.01
For lower than 10.000 viscosity cP: 0.1
For viscosity equal to or above 10.000 cP: 1
> Repeatability: 0.2%
> Thermometer features:
Temperature margins:
-40° C to + 300.0° C (- 40.0 ℉ to + 572.0 ℉ )
> Resolution: 0.1℃ / 0.1722 ℉
> Precision: +/- 0.1 ℃
> Type of probe: PT100
> Supplied at 100-240 VAC, 50/60 Hz
Spindles
AISI 316 stainless steel spindles, easily identified
by number and letter when changed according to range of viscosity.
These are the standards our units follow:
> BS: 6075, 5350
> ISO: 2555, 1652
> ASTM: 115, 789, 1076, 1084, 1286, 1417, 1439, 1638, 1824, 2196, 2336, 2364, 2393, 2556,
2669, 2849, 2983, 2994, 3232, 3236, 3716
Order Information